9.5 Error Correction
Most of the errors can be reset by power off/on. In the event of a short power
failure during a run, this run is interrupted and can be continued by pressing the
Start-key.
No indication on
the command panel:
Actions:
No power in the socket?
voltage present?
Input fuse ok?
Power switch on?
Contrast?
Check fuse in mains supply.
Plug in power cord correctly.
Switch on thermal fuse.
Switch on power.
Change contrast.
9.5.1 Centrifuge cannot be Started
a) Start-key LED not illuminated:
Close lid. Motorized lid locks must close.
b) Start-key LED illuminated:
Power off/on. If error occurs again, call
service.
9.5.2 Centrifuge Decelerates during Operation
Imbalance dialogue window is
active:
Rotor is loaded unevenly.
Centrifuge tilted.
Drive error (mechanical
damage).
Centrifuge was moved during
operation.
Centrifuge displays an error 73
to 77 after power on.
Balance loading.
Align centrifuge.
Call Service.
Restart after opening and closing lid.
These error numbers indicate an error in
the internal program storage. This error is
e.g. generated when a power failure
occurs during a storage procedure.