EasyManua.ls Logo

Sigma 4-15C - Installation of Rotors and Accessories

Sigma 4-15C
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Manual SIGMA 4-15C, page 37 of 95
03/06
3.3 Installation of Rotors and Accessories
1. Open centrifuge lid by pressing Lid-key.
2. Unscrew rotor tie-down screw from motor shaft (anticlockwise).
3. Lower the rotor straight down onto the motor shaft.
4. Tighten the tie-down screw (clockwise) with the rotor wrench so that the disc
spring is pressed together.
Fastening torque: 1-6/1-15/2-5/1-15K/2-16/2-16K/2-16KC: approx. 5 Nm
3-16/3-16K/3-18K/3K 30: approx. 7,5 Nm
4-15/4K15/6-15/6K15: approx. 10 Nm
In the event of frequent use the tie-down screw must be loosened by some
turns and fastened again. This should be done once a day or after approx.
20 cycles (please refer to chapter 5.1.4 Alteration of the Configuration
Cycles). This ensures a proper connection between rotor and shaft (please
refer to chapter 8.2 "Care and cleaning of accessories" as well).
5. Fill all positions of swing-out
rotors with buckets. Make
sure that all buckets are
inserted correctly.
6. Use only appropriate vessels
for the rotor (please refer to
chapter 1.2 "Suitable acces-
sories" as well).
Universal grease
allowed not allowed
7. Fill vessels external to the centrifuge.
8. Put or screw on covers of vessels.
9. Opposite places of the rotors must always be loaded with same accessories
and same filling.
10. Attention when using microtiter rotors: It is important not to run the plate
holders without plates inserted.
11. In angle rotors the plastic vessels must always be totally filled to avoid cracks
of vessels and leakages or loosening of the caps in case of partial filling.
Attention, follow the special comments of chapter 1.5.

Related product manuals