Operating Manual SIGMA 4-15C, page 43 of 95
03/06
4.2 Display
Default values
4.2.1 Set
If this area is illuminated, you are in the change mode which you can exit by
pressing the knob (here in combination with the speed).
4.2.2 Speed
Speed
In the upper section of the area the set speed of the centrifuge is displayed.
Underneath is the actual speed. The maximum speed values are rotor
dependent.
4.2.3 Relative Centrifugal Force (RCF)
RCF
The relative centrifugal force is the acceleration which the sample is exposed to.
The set value of this parameter is in the upper section of this area, underneath is
the actual value. The maximal RCF-values are rotor dependent.
(Refer to chapter 9.4 "Mathematical relations")