The Sinocare lancing device is prepared and ready to lance your finger for a blood
sample.
The lancet is for single use only. To prevent the risk of blood borne pathogen transmis-
sion, it shall not be shared with other people.
STEP 7:
Pull the spring cover back to arm the
lancing
STEP 8:
Take a strip out of the strip vial.
Immediately close the vial lid.
Attention: Finish test within 3 minutes after
take out strip. Otherwise test result may
be inaccurate.
STEP 9: IInsert strip into the test strip port.
For Safe-Accu 2, meter will turn on.
Tips: Insert the strip with the printed side facing
up and contact end into strip port.
Attention: If blood is not applied within 3 minutes,
the meter will automatically turn off. Reinsert the
strip and begin test.
If the meter display “E-2”, it means the environment temperature is beyond 40℃ or
below 5℃, Please put the meter, strip, control solution at appropriate place (10℃~35℃),
wait at least 30 minutes and test again.
STEP 10: Make sure the lancing device is pressed
firmly to your finger. Press the button to lance.
Attention: If you didn’t get enough blood sample,
please gently massage your finger. DO NOT
SQUEEZE YOUR FINGER. Refer to step 6 to
adjust puncture depth and lance again.