Introduction to the Your Blood Glucose Monitoring System
Intended Use
The Safe-Accu 2 Blood Glucose Monitoring System is designed for the quantitative
measurement of glucose in fresh capillary whole blood samples taken from the fingertip
and in venous whole blood samples. The Blood Glucose Monitoring System is for use
outside the body only (in vitro diagnostic use) for self-testing and professional use as an
aid in the management of diabetes.
The Safe-Accu 2 Blood Glucose Monitoring System is intended for in vitro diagnostic use
and should not be used for the diagnosis of or screening of diabetes.
The Safe-Accu 2 Blood Glucose Monitoring System includes:
Safe Safe-Accu 2 blood glucose meter, Safe-Accu 2 blood glucose test strip, blood
glucose control solution
NOTE: The blood glucose control solution is optional.
The Safe-Accu 2 meter display blood glucose results in mmol/L or mg/dL.
Each meter only displays one unit of measurement. The unit is preset and you can’t
change the setting. Please confirm the unit before testing.
Test Principle
A blood glucose test is based on measurement of electrical current caused by the
reaction of glucose with the reagents (special chemicals) on the electrode of the strip.
The blood or control solution sample is drawn into the tip of the test strip through capillary
action. Glucose in the sample reacts with the special chemicals and generates electrons,
which produce electrical current. The blood glucose meter measures the electrical current
and calculates the glucose result. The glucose result is displayed by your meter as mg/dL
or mmol/L.