To disinfect the meter
Please according to clinic institution disinfect method to disinfect the meter.
Power Saving
· If there is no blood applied to the test strip within 3 minutes, the meter will automatically
turn off.
· After test, the test results will be displayed on the screen with test strip in strip port, If
there no operation on the meter within 3 minutes, the meter will automatically turn off.
· If there is no other operation on the meter within 1 minute, such as enter into test
results review mode, the meter will automatically turn off.
Specifications
Blood volume 0.6μL
Sample type
Capillary whole blood, Venous whole blood
Calibration
Plasma equivalent
Measuring time
10s
Meter storage/transportation conditions
-20℃~55℃
Dimension
105*60*23(mm)
Weight
77.2g,with battery
Power source
3V DC,10mA,2 AAA alkaline batteries
Memory
200 blood glucose test results
Operating condition
Construction
Measurement units
Measurement range
10℃~35℃ ≤80% RH
Hand-held
mg/dL or mmol/L
20~600 mg/dL or 1.1~33.3 mmol/L
20141014(A.1.1)
Software Version(mg/dL)
20160112(A.2.1)Software Version(mmol/L)
Performance Characteristics
Accuracy: 95% test result of Safe-Accu 2 meet the requirement below:
Concentration Range
<5.5 mmol/L (100 mg/dL)
≥5.5 mmol/L (100 mg/dL)
Within±0.83 mmol/L(15 mg/dL)
Within ± 15%
Bias%
Precision: test result of Safe-Accu 2 meet the requirement below:
Details please read Safe-Accu 2 blood glucose test strip package insert.
Concentration Range
<5.5 mmol/L (100 mg/dL)
≥5.5 mmol/L (100 mg/dL)
SD< 0.34 mmol/L (6.0 mg/dL)
CV<6.0%
Requirement