XIV.Comparing Meter with Laboratory Results
When comparing results between Safe AQ SmartSafe AQ AirSafe AQ Voice
blood glucose monitoring system and a laboratory system, Safe AQ SmartSafe
AQ AirSafe AQ Voice blood glucose monitoring system blood tests should be
performed within 30 minutes of a laboratory test.
XIII.Information for Healthcare Professional Use
Only trained healthcare professional be allowed to use Safe AQ SmartSafe AQ
AirSafe AQ Voice blood glucose monitoring system on multiple patients.
Only trained healthcare professional be allowed to collect venous blood samples.
Any patient who has infectious diseases must use his/her own meter.
• If the same meter be used on multiple patients, there is risk of infection between
both patients and healthcare professional.
Water or other disinfection solution which remains on skin may dilute blood and
cause inaccurate test results.
Please dispose used lancet, test strips and expired control solution as medical
waste.
Perform test:
1. Wear disposable gloves.
2. Healthcare professional use an alcohol swab to clean lance site. Then be sure
dry stick site before testing.
3. Use qualified lancing device and choose suitable penetrate depth for patients.
4. Using lancing device to lance.
5. Hold the tip of the test strip to the blood sample until the test strip reaction zone
is full and the meter begin countdown.
6. Wash hands thoroughly with soap and water after handling the meter, lancing
device or test strips.
7. Cleaning and disinfecting meter and lancing device before use on next patient.