VIII.Pairing (Connecting) Meter with Smartphone (For Safe AQ Air only)
Pairing enables meter to upload test results when requested by an App (applica-
tion) on smartphone. The smartphone must have the bluetooth function turned on
and have an App that is able to receive the results.
Take examples of Wechat App.
1. Make sure that the Bluetooth Smart feature is turned on in the smartphone.
2. Open APP.
3. Scan QR code which provided with Safe AQ Air blood glucose monitoring
system.
4. Click “Link device”.
5. Insert test strip, meter turn on.
6. If the bluetooth symbol displayed, it means connect successful. Perform test
and your test result will be uploaded to your smartphone. If the bluetooth symbol
is flash, please check if you have turned on bluetooth function on your smart-
phone and insert test strip again.
Note: Different App may be little different while pairing meter with smartphone.
IX.Control Solution Testing
Why Perform a Control Solution Test
Ensures that your meter and test strips are working properly.
Allows you to practice testing without using blood.
When Should You Perform a Control Solution Test
When you begin using a new vial of Test Strips.
When the vial cap is opened for a long time.
When Test Strips have been exposed to extreme environmental conditions.
When you want to check whether meter or test strips are work properly.
When you drop or broken the meter.
When you want to check whether the test steps are correct.
Important Information
Use only Sinocare blood glucose control solution with your Meter.
Check the expiration date on the control solution bottle. Do not use if expired.
Use the test strips within a period of 6 months from the date that you first open
the vial.
Record the opened date on the control solution bottle, which is 3 months from
when you first open it. Discard the bottle after 3 months.