Precision: test result of Safe AQ Smart, Safe AQ Air, Safe AQ Voice meet the requirement below:
Details please read Safe AQ blood glucose test strip package insert.
XVIII.Specications
29
30
Concentration Range
<5.55 mmol/L (100 mg/dL) SD< 0.34 mmol/L (6.0 mg/dL)
CV<6.0%≥5.55 mmol/L (100 mg/dL)
Requirement
Blood volume 0.6μL
Sample type
Capillary whole blood, Venous whole blood
Calibration
Plasma equivalent
Measuring time
Meter storage/transportation conditions
Dimension
Weight
Power source
Software Version A
-20℃~55℃
84*60*26(mm)
73.5g with battery
3V DC, 50mA, 2 AAA alkaline batteries
5s
Memory 10 blood glucose test results
Testing condition
10℃~35℃ ≤80% RH
Construction
Hand-held
Measurement units
Measurement range
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time count-down
Temperature Warning
Battery power Warning
Used test strips Warning
20~600 mg/dL or 1.1~33.3 mmol/L
Shelf life
10 years(estimated by test 7 times per day).
During use, user should maintain the product
refer to this user manual’s requirements.
Additional features
mg/dL or mmol/L