7
8
FOR Safe AQ Voice meter:
1. If test results greater than 600 mg/dL(33.3mmol/L), the meter will display
“HI” and read“The result is beyond the range”with red light on. Retest your blood
glucose immediately with a new test strip. If your reading is still “HI”, contact
healthcare professional immediately.
2. If test results is lower than 20 mg/dL(1.1mmol/L), the meter will display “LO”
and read “The result is below the range”with red light on. Retest your blood
glucose immediately with a new test strip. If your reading is still “LO”, contact
healthcare professional immediately.
3. If test result is >70mg/dL(3.9 mmol/L)and < 126 mg/dL(< 7.0 mmol/L),
the meter will display and read the test result with green light on.
4. If test result is ≥ 126 mg/dL(≥7.0 mmol/L)and ≤ 250 mg/dL(≤13.9 mmol/L),
the meter will display and read the test result with yellow light on.
5. If test result is ≤ 70mg/dL(≤ 3.9 mmol/L)or >250mg/dL(> 13.9 mmol/L),
the meter will display and read the test result with red light on.
Note: Do not change your treatment only based on test result and light indications.
Please consult healthcare professional for your treatment.
V.About Your Safe AQ Smart、Safe AQ Air、Safe AQ Voice Meter
Test strip port
Strip eject button
LCD display
Main button
Test strip port
Strip eject button
LCD display
Main button
Battery Cover
Battery Cover
Safe AQ Smart Meter:
Safe AQ Air Meter: