XV.Cleaning and Disinfecting
Clean and disinfect meter immediately after getting any blood on the meter or if
meter is dirty.
• If the meter is being used on second person, the meter and lancing device
should be clean and disinfected first. Do not clean the meter during the test.
To clean the meter:
1. Wash hands thoroughly with soap and water.
2. Make sure meter is off and a test strip is not inserted. Use 75% alcohol to rub
the entire outside of the meter. Make sure no liquids enter the test strip port or
other opening in the meter.
3. Let meter air dry thoroughly before using to test.
4. Wash hands thoroughly again after handing meter.
5. Verify the meter is working properly by performing control solution test. If your
meter can’t work properly, please contact customer service at +86-731-89935581/
+86-731-89935582 or contact local dealers.
To disinfect the meter:
Please according to clinic institution disinfect method to disinfect the meter.
XVI.Power Saving
• If there is no blood applied to the test strip within 3 minutes, the meter will
automatically turn off.
After test, the test results will be displayed on the screen with test strip in strip
port, If there no operation on the meter within 3 minutes, the meter will automati-
cally turn off.
Note: If test result display “LO” or “HI”, the test results will be displayed 1 minute.
If there is no other operation on the meter within 1 minute, the meter will automati-
cally turn off.
If there is no other operation on the meter within 1 minute, such as enter into test
results review mode, the meter will automatically turn off.
XVII.Performance Characteristics
Accuracy: 95% test result of Safe AQ Smart, Safe AQ Air, Safe AQ Voice meet the requirement below:
27
28
Concentration Range
<5.55 mmol/L (100 mg/dL) Within±0.83 mmol/L(15 mg/dL)
Within ± 15%≥5.55 mmol/L (100 mg/dL)
Bias%