Page 32
ERGON SERVICE • REV0
TYPICAL DOCK-IT PUMP
WIRING DIAGRAM
POWERED HEIGHT
ADJUSTABLE ARM
PUMP ENCLOSURE (18"L x 12"H x 6"D)
(REMOTE MOUNTED)
3/4" CONDUIT AND MINIMUM 12 AWG WIRE SIZE
(2 WIRES PER LIGHTHEAD PLUS FIXTURE
GROUND WIRE) REQUIRED BETWEEN FIXTURE
JUNCTION BOX AND SKYTRON SUPPLIED
WALL CONTROL. PLEASE REFER TO SPECIFIC
MODEL LIGHT WIRING DIAGRAM PROVIDED.
SURGICAL LIGHTS JUNCTION BOX (4"L x 4"W X 2-1/4"D)
(FIXTURE MOUNTED)
GROUP 3
GREEN
TO SKYBOOM
FIXTURE
ORANGE/BLUE (N)
BROWN/YELLOW (L)
GREEN (G)
PUMP DRAW:
2.7 AMP
120VAC, 60HZ, 1PH
12AWG, XHHW-2
CIRCUIT CONDUITS
PROVIDED BY OTHERS
120 VAC, 20 AMP
GROUP 2
120 VAC, 20 AMP
GROUP 1
120 VAC, 20 AMP
FACILITY
SUPPLIED
YELLOW
FIXTURE
TERMINAL
STRIP
7
6
5
4 2 1
3
GREEN
24V
SWITCH
CIRCUIT
CONNECTOR
43
2
1
3
1
2
BLACK
WHITE
WHITE
BROWN
NC
CIRCUIT BOARD
GREY
YELLOW
CIRCUIT BREAKER
120VAC, 10 AMP
UP
LIMIT
SWITCH
GREEN/YELLOW
PUMP
1
PUMP HOUSING
TERMINAL STRIP
DOWN
LIMIT
SWITCH
FACILITY INCOMING POWER
SUPPLY LINES MUST BE PROPERLY
CIRCUIT PROTECTED AND IN
COMPLIANCE WITH FEDERAL,
STATE, AND LOCAL CODES.
NC
8
HANDLE
CONTR OL
PENDANT
CONTR OL
WHITE
BROWN
GREEN
YELLOW
BROWN
GREEN
YELLOW
GREYGREY
YELLOW
GREEN
BROWN
WHITE
BROWN
BLUE
BROWN
BLUE
WHITE
WHITE
BROWN
GREEN
YELLOW
GREY
BROWN
BLUE
BROWN
BLUE
GREEN
GREY
BROWN
WHITE
BLUE
BROWN
BLACK
2
3
4
5
6
BROWN
BROWN
GREEN/YELLOW
GREEN/YELLOW
BLUE
BLUE
BROWN
BLUE
BLACK
BLUE
BROWN
GREEN/YELLOW
WHITE
GREEN
CART ENGAGEMENT
LEFT
CART ENGAGEMENT
RIGHT
WHITE
BROWN
GREEN
YELLOW
PCB
WHITE
SOLENOID
THERMO-
COUPLE
MOTOR
START
CAPACITOR
BLACK
WHITE
TYPICAL DOCK-IT PUMP WIRING DIAGRAM