10011896-001
13
Removing a Cassette
WARNING: Close the fluid path tubing with the clamp
before removing the cassette from the pump to prevent
unregulated gravity infusion, which could result in death
or serious injury.
To remove a cassette:
1. Press and hold
.
2. When you see 3 sets of
dashes appear, release
.
3. Close all tubing clamps and
disconnect the tubing from your access
device as instructed by your clinician.
4. Insert a coin into the slot in the lock and
turn it clockwise. The lock will pop out
when you unlock the cassette.
5. A continuous alarm will sound and the pump will display No
Disposable, Clamp Tubing. The alarm may be silenced by
pressing
or
.
6. Remove the cassette from the pump
hinge pins.
7. Discard the CADD
medication
cassette reservoir or CADD
®
administration set as instructed by
your clinician.
Stopping
---- ---- ----
10011896-001.indd 13 2/27/2018 7:08:13 PM