4
Cautions
Do not operate the pump at temperatures below +2°C (36°F)
or above 40°C (104°F).
Do not store the pump at temperatures below -20°C (-F)
or above 60°C (140°F). Do not store the pump with a CADD
medication cassette reservoir or CADD
®
administration set
attached. Use the protective cassette provided.
Do not expose the pump to humidity levels below 20% or
above 90% relative humidity.
Do not store the pump for prolonged periods with the
batteries installed.
Frozen medications must be thawed at room temperature
only. Do not heat the CADD
medication cassette reservoir
in a microwave oven as this may damage the medication, the
CADD
medication cassette reservoir, or cause leakage.
Do not immerse the pump in cleaning fluid or water or allow
solution to soak into the pump, accumulate on the keypad,
or enter the battery compartment.
Do not clean the pump with acetone, other plastic solvents,
or abrasive cleaners.
Do not use the pump in the presence of flammable
anesthetics or explosive gases.
Use only Smiths Medical accessories as using other brands
may adversely affect the operation of the pump.
10011896-001.indd 4 2/27/2018 7:08:09 PM

Other manuals for Smiths Medical CADD-legacy 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-legacy 1 and is the answer not in the manual?

Smiths Medical CADD-legacy 1 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-legacy 1
CategoryMedical Equipment
LanguageEnglish

Related product manuals