10011896-001
18
WARNING: Ensure that the entire fluid path is free of all
air bubbles before connecting to your catheter to prevent
air embolism. Air embolism could result in death or serious
injury.
Inserting the Tubing into the Air Detector
The Air Detector is designed to detect air bubbles in the fluid
path. When the tubing is routed through an Air Detector that is
turned on, an air bubble exceeding the specified size will cause
an alarm to sound and the pump to stop.
1. If your clinician instructed you to use the Air Detector, make
a small loop of tubing underneath the air detector and hold
it with your thumb.
2. Place the tubing over the groove in the air detector and tuck
it under the catch.
Catch
10011896-001.indd 18 2/27/2018 7:08:15 PM