22
Messages and Alarms
When you see and hear: Take this action:
Two-Tone alarm
High Pressure
There may be a kink in the
tubing or a clamp may be
closed. Unkink the tubing
or open the clamp and the
pump will resume delivery.
You may press
or
to stop the pump and silence
the alarm for 2 minutes.
After you remove the cause
of high pressure, start the
pump if necessary. If the
alarm continues, contact your
clinician.
Two-Tone alarm
Key pressed,
Please release
If a key is being pressed,
stop pressing it. If the alarm
persists, close the tubing
clamp and remove the pump
from use. Contact your
clinician.
Three Two-Tone beeps
every 5 minuTes
LowBat
The AA batteries are low
but the pump is operable.
Change the AA batteries
soon.
Two-Tone alarm
Motor Locked,
remove all power
Batteries are depleted and
the pump was powered up
with the AC Adapter. Install
new AA batteries, reconnect
the AC adapter, and restart
the pump.
10011896-001.indd 22 2/27/2018 7:08:15 PM

Other manuals for Smiths Medical CADD-legacy 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-legacy 1 and is the answer not in the manual?

Smiths Medical CADD-legacy 1 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-legacy 1
CategoryMedical Equipment
LanguageEnglish

Related product manuals