EasyManua.ls Logo

Snap-On Diagnostics DGA 1000 - Sample System O2 and NO Cell Service

Snap-On Diagnostics DGA 1000
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
4-3
E Primary Filter Bowl Assembly
Removes moisture and other exhaust contaminants that are
potentially harmful to the analyzer.
Inspect daily and replace every two week with part number
7096E9062-77.
To replace the filter, unscrew the locking ring, which
connects the bowl to the upper housing, to gain access
to the filter.
Unscrew the filter retainer nut on the bottom, and
remove and install the new filter.
When the filter bowl is removed, be careful not to loosen
or misalign the o-ring seal in the top portion of the filter
housing. If the o-ring does not properly seal, a failed leak
check will occur.
If desired and also recommended, the bowl may be
washed with soap and water. Never
use a solvent on the
bowls.
Do not overtighten filter retainer or bowl ring. Snug is fine.
If the water symbol ( ) appears on the left of the
screen or if a low flow symbol ( ) appear at any time
during operation, the primary filter must
be replaced
immediately.
Stop testing and perform sample system
maintenance.
The low flow ( ) can also be caused by a
plugged sample hose or probe. Check these along
with sample system service.
P Calibration Gas Inlet
Use this inlet to supply calibration gas to the analyzer.
Pressure at this port must not exceed 10 psi.
S Clean Air Inlet Port, Charcoal Filter
A charcoal filter is installed to filter the ambient air going
into the gas bench during calibration.
Replace yearly with part number 7096E9061-54.
To replace the filter, remove the old filter from the tubing
and insert the new filter.
If one is present, pay attention to the flow arrow.
O
2
Cell and NO Cell Service
Do not attempt to replace these cells on your own. When the
warning icon ( ) displays in the lower right of the screen, cell
replacement is required. Contact
Equiserv
.
Under most conditions, depending on use, the NO cell
lasts 1518 months and the O
2
cell 1218 months.

Table of Contents