4-2
Maintenance
Sample System
The analyzer contains three filters and a charcoal filter to protect
and extend the analyzer sample system life.
Three filters, A, C and E shown in
Figure 4-1
, must be serviced at
regular intervals. How often depends to a large extent on the use
of the analyzer. The following are minimum intervals for
replacement.
AFilter, Secondary Inline
Sample, secondary in-line filter.
Replace every six months with part number 7096E9061-03.
To replace the filter, remove the tubing from each end
and insert the new filter.
Pay attention to the flow arrow.
CFilter, Water Filter
In-line water drain filter.
Replace every six months with part number 7009E9323-46.
To replace the filter, remove the tubing from each end
and insert the new filter.
Pay attention to the flow arrow.

:
AnalyzerBack Panel View