Maintenance
4-5
4O
2
/Check Install
This function must be used when the O
2
is replaced. It is not
necessary to use this function at any time other than O
2
cell
replacement.
5Gas Tag Values
Use the Gas Tag Values function to enter calibration gas bottle
values of the calibration gas used for gas calibration.
If displays next to the menu choice, no entry can be
made.
If values are entered, the following displays:
Type 1 Low, used to enter low range cal gas for:
CO 1.0
CO
2
6.0
Propane 300
Values shown here are recommended.
Type 2 High, used to enter high range cal gas for:
CO 8.00
CO
2
12.0
Propane 3200
Values shown here are recommended.
Type 3 Low, used to enter low range NO if required:
NO 300
Type 4 High, used to enter high range NO if required:
NO 3000
Only
Equiserv
can perform a two point calibration.
6Gas Calibration
As delivered, the analyzer can be gas calibrated by a Service
Technician with one or two cylinders of calibration gas. If you
desire to perform the calibration, use the gas regulator, part
number 0131-0031-01, and hose, part number, 5-052351-405 for
the procedure.
It is recommended that
Equiserv
calibrate your analyzer
once every six months. However, the following procedure
will allow you to calibrate the analyzer if desired.
The recommended calibration gas, part number
0271-0099-01, is gas with concentrations of:
CO 8.0
HC 3200 = propane
CO
2
12.0
NO 3000
Without the NO option, part number 0271-0088-01
CO 8.0
HC 3200 = propane
CO
2
12.0