Do you have a question about the Soredex Cranex D and is the answer not in the manual?
Ensure the main pan unit, ceph arm, and head are level before starting the alignment.
Guide to enter the unit's service mode using specific key presses and selections.
Instructions for installing the beam alignment tool onto the Ceph sensor.
Procedure to remove the secondary collimator from the Ceph head by unplugging the ribbon cable.
Configure the unit to 85 kV, 10 mA, and 2.00 seconds for beam visibility.
Perform an exposure to check beam alignment against tool lines.
Loosen screws and slide head to center beam horizontally using the alignment tool.
Adjust collimator bolts and alignment screws for vertical beam centering on the tool.
Reattach the secondary collimator and ribbon cable after beam centering.
Verify beam alignment on the tool, redo steps 7-9 if necessary.
Perform advanced adjustments for horizontal, vertical, and tilt alignment.
Loosen cog wheel screw and turn to align the belt and secondary properly.
Loosen mounting screws and adjust top screw for vertical beam position.
Adjust the side screw to correct diagonal beam direction.
Adjust trimpots for sensitivity and perform a scan to confirm operation.
Turn trimpots counter-clockwise to increase board sensitivity for beam detection.
Run a Ceph scan in normal mode to ensure it completes fully.
Prepare for ear rod alignment by checking current position of rings.
Loosen bolt and adjust screw for vertical ear rod positioning.
Loosen rear screw and adjust rod for horizontal ear rod positioning.
Perform final exposure, confirm scan and ear rod settings. Unit is done.
This document outlines the procedure for Cranex® D Site Ceph Beam Alignment, specifically for the White Panel Version of the Cranex D Ceph unit. It serves as a quick guide for technicians to perform the beam alignment, emphasizing the use of a dedicated beam alignment tool (Part # 203603).
The primary function of this procedure is to ensure the precise alignment of the X-ray beam in the Cranex D Ceph unit. Proper beam alignment is crucial for accurate cephalometric imaging, which is used in orthodontics and oral surgery for diagnosis, treatment planning, and monitoring. The alignment process involves adjusting both horizontal and vertical aspects of the beam, as well as the positioning of the Ceph head and ear rods, to ensure the X-ray beam is centered and correctly positioned relative to the imaging sensor and patient positioning aids.
The procedure is broken down into several steps, guiding the user through a systematic alignment process:
This document itself is a maintenance procedure, providing step-by-step instructions for calibrating the X-ray beam. It highlights specific components that require adjustment during maintenance, such as:
The procedure emphasizes the importance of a dedicated beam alignment tool (Part # 203603) for accurate and efficient maintenance. It also implicitly suggests the need for a trained technician to perform these adjustments due to the specific tools and knowledge required for precise calibration of medical imaging equipment. Regular beam alignment ensures consistent image quality and patient safety.
| Brand | Soredex |
|---|---|
| Model | Cranex D |
| Category | Dental equipment |
| Language | English |
Loading...