14 ENGLISH
6. SPECIAL MODES
6.1. NOMINAL MODE (EMERGENCY MODE)
A safety function may be rapidly restored by pressing the following
button:
In nominal mode, the pacemaker operates with the parameters
underlined in the table of programmable parameters.
6.2. STANDBY MODE
In the event of self-test failure, the pacemaker switches
automatically to a safe mode of operation known as standby mode.
In most cases, normal pacemaker operation may be re-established (if
the pacemaker is at a stabilized temperature of approximately 37°C)
by validating the reinitialization function suggested by the
programmer. If this is not the case, contact your Sorin CRM
representative.
In standby mode, the pacemaker operates with the following
parameters: VVI, 70 min
-1
, 5 V, 0.5 ms, sensitivity 2.2 mV, unipolar.
Programming is then impossible and test modes, including magnet
mode, are ineffective.