ENGLISH – 35
10.1. CURRENT DRAIN AT BEGINNING OF LIFE
Shipment conditions (1) safeR mode (3) Inhibited: 6.5 µA
100% pacing: 10.4 µA
Shipment conditions (1) DDD mode Inhibited: 6.5 µA
100% pacing: 14.2 µA
Cenelec conditions (2) safeR mode (3) Inhibited: 6.9 µA
100% pacing: 9.6 µA
Cenelec conditions (2) DDD mode Inhibited: 6.9 µA
100% pacing: 12.3 µA
(1) 60 min
-1
, 3.5 V, 0.35 ms, 500 Ohms, Holter ON, sensors ON.
(2) 70 min
-1
, 2.5 V, 0.5 ms, 500 Ohms, Holter ON, sensors ON.
(3) Twenty minutes after the detection of implantation, the device
switches automatically from DDD to safeR mode, assuming 1%
ventricular pacing and 100% atrial pacing.
10.2. CORRESPONDENCE BETWEEN MAGNET
RATE AND BATTERY IMPEDANCE
Magnet rate (min
-1
) 96 95 94 93 91 90
Magnet period (ms) 625 633 641 648 656 664
Battery resistance (kilohms) <3 3 4 5 6 7
Magnet rate (min
-1
) 89 88 86 83 82 80
Magnet period (ms) 672 680 695 719 734 750
Battery resistance (kilohms) 7.5 8 8.5 9 9.5 10
Magnet rate (min
-1
) 78 75 73 70
Magnet period (ms) 773 797 820 859
Battery resistance (kilohms) 11 12 13 >14