ENGLISH – 37
11. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold: "as shipped" value
Underlined value
: nominal value
Basic parameters Values
Mode
(1)(2)
DDD-DDDR-DDD/DDIR-safeR-
safeRR-safeR/DDIR-Dplus-Dplus-
R-Dplus/DDIR-DDTA-DDTV-
DDTAV-VDD-VDDR-DDI-DDIR-
DOO-VVI
-VVIR-VVT-VOO-AAI-
AAIR-AAT-AOO-OOO
Basic rate (min-1)
(3)
30-40-45-50-55-60-65-70
-75-80-
85-90-95 (± 2 min-1)
Rest rate (min-1)
50-55-60-65-70-75-80-85-90-95
(± 2 min-1)
Maximum tracking rate (min-1)
100-110-120-130-140-155-165-
175-185 (± 5 min-1)
Rate hysteresis (%) 0-5-10-20-35 (± 4 %)
Rest AV delay (ms)
30-45-65-80-95-110-125-140-155-
170-190-205-220-235-250 (± 10 ms)
Exercise AV delay (ms)
30-45-65-80-95-110-125-140-155-
170-190-205-220-235-250 (± 10 ms)
AVD Paced/Sensed Offset (ms) 0-15-30-45-65-80-95-110-125 (± 5 ms)
(1) Twenty minutes after the automatic detection of implantation by the device, the
as-shipped pacing mode (DDD) is automatically reprogrammed to SafeR.
(2) Do not program OOO mode in pacing-dependent patients.
(3) The corresponding basic periods and escape intervals are as follows:
1961-1500-1333-1200-1091-1000-923-857-800-750-706-667-632 ms.