38 – ENGLISH
Pacing/Sensing Values
Atrial or ventricular amplitude (V)
(1)
1.5-2.0-2.5-3.0-3.5-4.0-5.0
-7.5
(± 20 %)
Atrial or ventricular pulse width
(ms)
(1)
0.10-0.25-0.35-0.50
-0.60-0.75-
0.85-1 (± 35 µs)
Atrial sensitivity (mV)
(2)(3)(4)
0.1-0.2-0.3-0.4 (± 0.025 mV ±
40 %)
0.6-0.8-1.0-1.2-1.5-1.8-2.0-2.2-2.5-
2.7-3.0-3.5-4.0-4.5-5.0-6.0
(± 20 %)
Ventricular sensitivity (mV)
(2)(3)
1.0-1.2-1.5-1.8-2.0-2.2
-2.5-2.7-3.0-
3.5-4.0-4.5-5.0-6.0-8.0-10.0-15.0
(± 20 %)
Atrial or ventricular sensing polarity
(5)
Unipolar
-Bipolar
Atrial or ventricular pacing polarity
(5)
Unipolar
-Bipolar
(1) The nominal value may be greater than the underlined value (return to
programmed value).
(2) Values are measured using a positive and negative triangular signal of 2/13
ms.
(3) For sensitivity settings below 2.5 mV, the pacemaker may detect noise lower
than the level specified in clause 27.5.1 of standard EN 45502-2-1.
(4) For sensitivity settings below 0.3 mV, the pacemaker may detect noise
lower than the level specified in clause 27.4 of standard EN 45502-2-1
(5) As soon as the detection of implantation has been confirmed, the lead
configuration is automatically programmed to unipolar pacing and
bipolar sensing (if a bipolar lead is used).