ENGLISH 39
Special features
Values
Smoothing OFF-Very slow-Slow-Medium-Fast
Acceleration (%) 0-5-15-25-35-45
AV delay shortening (ms) 0-15-30-45-65-80-95-110
Mode switching ON-OFF
Anti-PMT protection Termin-Reprog
Atrial or ventricular Autosensing Auto-Monitor
Ventricular Auto-threshold Auto-Monitor-OFF
Minimum ventricular amplitude (V) 1.5-2-2.5-3.0-3.5 (± 10 %)
Post V Atrial Blanking (ms)
(1)
150-165-180-195-210-225-240-255
(± 10 ms)
(1) After ventricular pacing, the minimal blanking in atrium is equal to the
programmed value. After ventricular sensing, the minimal blanking in
atrium is equal to the programmed value minus 55 ms.
AA prevention functions Values
Pause suppression A-V-A+V-OFF
Acceleration on PAC ON-OFF
Overdrive ON-OFF
Maximum Overdrive rate (min-1) 100-110-130-155-185 (± 15 min-1)