EMC Precautions
Special precautions are required regarding the Electromagnetic Compatibility (EMC) of
the CVX-300. The CVX-300 needs to be installed and put into service according the EMC
information provided in this manual.
Portable and Mobile Radio Frequency (RF) communications equipment can aect any
medical electrical equipment including the CVX-300.
Only cables and accessories provided by Spectranetics may be used with the CVX-300. The
use of any other cable or accessories may have an adverse eect on the Electromagnetic
Compatibility of the CVX-300 such as increased emissions or decreased immunity.
The CVX-300 should not be used adjacent to or stacked with other equipment. Shoulduse
adjacent to other equipment become necessary, the CVX-300 should be observed to verify
normal operation in that conguration.
Table 201 (EN 60601‑1‑2)
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions
The CVX‑300 is intended for use in the electromagnetic environment specied
below. The customer or the user of the CVX‑300 should assure that it is used in
such an environment.
Emissions Test Compliance Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The CVX‑300 uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The CVX‑300 is suitable for use in all
establishments other than domestic
and those directly connected to the
public low‑voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000‑3‑2
Class A
Voltage Fluctuations/
icker emissions
IEC 61000‑3‑3
Complies