Fault: If the Fault advisory light illuminates with a fault code in the display window, it
indicates a malfunction. Record the fault code and notify Spectranetics for service.
Fiber: The Fiber advisory light indicates the ber is not connected or properly inserted into
the ber adapter.
Fluence and mJ/mm
2
: The Fluence and mJ/mm
2
indicator lights illuminate simultaneously
when the uence setting is visible in the display window.
Footswitch: The Footswitch advisory light indicates the footswitch is not connected, not
operating properly or is depressed during the Warm‑up or Standby mode.
Lasing: The Lasing status light indicates the footswitch is being depressed and the Excimer
Laser System is lasing. The red status light on the top of the control panel also illuminates
when the system is lasing.
Min‑Sec: The Min‑Sec indicator light is illuminated when the warm-up time or treatment
time is visible in the display window.
Power Error: The Power Error advisory light indicates the laser energy is out of range and the
Safety Power Monitor has disabled the laser beam. An audible tone will be heard whenever
the Power Error lamp is illuminated. During normal operation, this light may icker as the
laser energy varies within the acceptable range. The CVX-300 will continue to lase as the
unit is working to correct the condition. The shutter will remain closed while Power Error
is illuminated, preventing energy delivery through the ber. Refer to the Troubleshooting
section in this manual. If the Power Error and yellow advisory light on top of the control
panel illuminate constantly, this indicates a problem with the Excimer Laser System. Call
Spectranetics Customer Service for assistance.
Pulses: The Pulses indicator light illuminates when the Pulses Delivered button is
depressed and the number of laser pulses delivered is visible in the display window.
Rate and Pulses/Second: The Rate and Pulses/Second indicator lights illuminate
simultaneously when the laser pulse repetition rate is visible in the display window.
Ready: The Ready button illuminates indicating the Excimer Laser System is operating in
Ready mode. The yellow status light on top of the control panel also illuminates when the
system is in Ready mode.
Service: The Service advisory light indicates that the laser is near its maximum energy
output and requires service. The Excimer Laser System is still operational and safe. Call
Spectranetics immediately to schedule service.