Power error indicator
is illuminated constantly.
Call Spectranetics
Customer Service
for assistance.
A fault occurs during Warm‑up
or Calibration, the fault indicator
is illuminated and a fault code
number is shown in the middle
of the display window.
Refer to the Fault Code Table
at the end of this section.
The laser gives an odor
of HCl gas (like bleach).
Place the laser in a
well‑ventilated, yet isolated
room. Call Spectranetics
Customer Service to schedule
an emergency service visit.