Page 12
it is correctly positioned
Ensure waveform quality (use the guidance
bar).
After completing the above steps, repeat
PWV measurement
Unable to generate PWA Report
(Code= 3053).
Ensure use of correct cuff size, correct
positioning and fit
Ensure waveform quality (use the guidance
bar).
After completing the above steps, repeat
PWA measurement.
Unable to generate PWA Report
(Code= 3054).
Ensure there is enough memory in the
computer.
Ensure that the user has read/write
permission where the XCEL software
installed.
Ensure use of correct cuff size, correct
positioning and fit
Ensure waveform quality (use the guidance
bar).
After completing the above steps, repeat
PWA measurement.
Unable to start Capture (Code=
3060).
Ensure that the device is connected to the
computer.
Ensure that the cuff is connected to the
device.
Ensure use of correct cuff size, correct
positioning and fit
Turn the device OFF and ON.
After completing the above steps, attempt
Capture again.
Unable to complete the operation
due to system error (Code=3101)
Close SphygmoCor XCEL software
Turn the device OFF and ON.
Open SphygmoCor XCEL software
Unable to calculate PWV (Code=
3125).
Ensure that the entered distance in mm is
correct.