Page 11
(a) Unable to find the
Electronics Module due to
mismatch serial number
(b) Unable to start Brachial
calibration due to
mismatch serial number
(c) Unable to start Capture
due to mismatch serial
number
One of these messages related
to “incomplete calibration”
error will appear:
(a) Unable to find the
Electronics Module due to
incomplete calibration
(b) Unable to start Brachial
calibration due to
incomplete calibration
(c) Unable to start Capture
due to incomplete
calibration
Close SphygmoCor XCEL Software
XCEL Electronics Module need to be calibrated.
Please refer to SphygmoCor XCEL Calibration
KIT or contact AtCor Medical for technical
support.
Unable to generate PWV Report
(Code= 3051).
Ensure that the entered distance in mm is
correct.
Ensure use of the correct cuff size and that
it is correctly positioned
Ensure waveform quality (use the guidance
bar).
After completing the above steps, repeat
PWV measurement
Unable to generate PWV Report
(Code= 3052).
Ensure there is enough memory in the
computer.
Ensure the user has read/write permission
for the computer on which the SphygmoCor
XCEL software installed
Ensure the entered distance in mm is
correct.
Ensure use of the correct cuff size and that