Page 10
select Find Module.
Attempt Capture again.
Unable to continue Capture due to
Communication error
(Code=2077)
(Code=2078)
Press the STOP button on the device.
Turn the device OFF and ON.
Ensure that the device is connected to the
computer.
Exit the XCEL software
Start the XCEL software
In the Setup screen, click on the System
and select Find Module.
Attempt Capture again.
Unable to continue Capture due to
deflation of cuff (Code = 2079).
Ensure that the patient’s limb with cuff is
stationary.
Ensure use of the correct cuff size and that
it is correctly positioned
Ensure that there are no obstructions
between the cuff and arm.
After completing the above steps, attempt
Capture again.
Unable to continue Brachial
Calibration due to brachial pressure
being outside the normal range
(Code=2080)
Ensure systolic pressure is within the range
50 to 260 mmHg and diastolic pressure is
within the range 40 to 200 mmHg.
Ensure use of correct cuff size, correct
positioning and fit
Ensure that there are no obstructions
between the cuff and arm.
After completing the above steps, repeat
measurement.
Unable to continue Brachial
Calibration due to timeout.
Turn the module OFF/ON
Click OK button
Proceed with the measurement
One of these messages related
to “mismatch serial number”
error will appear:
In the Set Up screen, click System Key
Enter the correct system key provided by AtCor
Medical