Page 9
Error Code
Error Message
Action
the cuff
Ensure use of the correct cuff size and that
it is correctly positioned
Turn the device OFF and ON.
Attempt Capture again.
2049
2050
Unable to continue Capture due to
unsuccessful inflation of cuff (Code
= 2049).
(Code = 2050)
Ensure the device is connected to the
computer.
Ensure the cuff is connected to the device.
Ensure use of the correct cuff size and that
it is correctly positioned
Turn the module OFF and ON.
Attempt Capture again.
2051
2053
Unable to continue Capture due to
unsuccessful deflation of cuff (Code
= 2051)
(Code = 2053)
Press the STOP button on the device to
deflate the cuff.
Ensure that the device is connected to the
computer.
Ensure that the cuff is connected to the
device.
Ensure use of the correct cuff size and that
it is correctly positioned
Turn the device OFF and ON.
Attempt Capture again
2058
Unable to continue Capture due to
exceeding the maximum number of
inflation/deflation cycle per
measurement (Code = 2058).
Ensure that the patient’s limb with cuff is
stationary during measurement.
Ensure use of correct cuff size, correct
positioning and fit.
Ensure that there are no obstructions such
as thick clothing between the cuff and arm.
For PWV mode: Ensure the tonometer is
detecting pulses.
After completing the above steps, attempt
Capture again.
2076
Unable to continue Capture due to
invalid data from the Electronics
Module (Code = 2076).
Press the STOP button on the device.
Ensure that the device is connected to the
computer.
In the Setup screen, click on the System and