EasyManua.ls Logo

SphygmoCor XCEL - Page 40

SphygmoCor XCEL
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 40
Note: Once pumping starts ensure to click Continue within 3 minutes. If not, a message will appear
informing that the calibration was unsuccessful. Then close the SphygmoCor XCEL Calibration Software,
turn off/on the module and repeat the calibration procedures from Step 1. Please refer to Restarting the
Calibration Procedure Section in the Calibration Troubleshooting section.
27- The software will display this message.
Note: If this message appears (calibration was unsuccessful), then turn OFF and ON the device and close
the SphygmoCor® XCEL Calibration Software and repeat the calibration procedures from step 1. Please
refer to Restarting the Calibration Procedure Section in the Calibration Troubleshooting section.