EasyManua.ls Logo

Starrett IQ - Technical Specifications

Starrett IQ
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Specications
Measurement range ..250” / 6mm
.600” / 15mm
1.00” / 25 mm
Max error (.000050” /
.001mm Resolution)
.0001” / .002mm .0001” / .002mm .0001” / .002mm
Max error (.0001” /
.002mm Resolution)
.0002” / .004mm .0002” / .004mm .0002” / .004mm
Repeatability .000050” / .0012mm .000050” / .0012mm .0001” / .002mm
Weight .41lb (.18kg)
Measurement force 150 gm (1.47N) 150 gm (1.47N) 155 gm (1.52N)
Power battery 2 x 3V lithium battery CR2450
Data Cable 5V
Average battery life 6,000 hours (ave. life decreases using TIR, MAX, MIN, Radio, TOL
LEDs)
Data output USB, Digimatic (MTI)
Cables USB, 10 pin Digimatic (MTI) , RS232, Pigtail
Working temperature
(storage)
5
o
C to 40
o
C ( -10
o
C to 60
o
C)
40
o
F to 105
o
F ( -15
o
F to 135
o
F)
Error Cause Notes
Error 3 Failed to load saved settings Contact support if issue persists
Error 4 Value is too large to display Zero gage or reduce spindle travel
Error 5 Failed to load factory conguration Device may not operate correctly, contact support
Error 10 Incomplete lookup table Please download new lookup table or disable function
Error Codes