5-2
P093932-203 Equipment Manual Operator Troubleshooting
Table 5-1. Operator Troubleshooting
Problem Solution
1. The lighthead does not remain in position. 1. For ceiling mount, check the leveling of the pipe
anchored to the ceiling. For mobile floor stand,
check the leveling of the shank to the base.
2. Tighten the brakes to the articulations in order to
increase the friction.
2. The lighthead does not work. 1. Check the presence of the fuses inside the
terminal box (T1A), the hooking of the electrical
connectors.
2. Check the presence of voltage in the lighthead.
3. The fuse keeps burning. Check the characteristics of the inserted fuses.
4. The lighthead flickers. Contact STERIS.
5. The light beam on the examination field is not in
focus.
Contact STERIS.
6. The lighthead does not turn on. 1. Check the power supply voltage and fuses.
2. Electronics are damaged: contact STERIS.