8-1
EMC Compliance Technical Data Equipment Manual P093932-203
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions for all Equipment and
Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity – for all Equipment and
Systems (per IEC 60601-1-2 clause 6.8.3.201 a – 6 Table 202)
Table 8-1. Guidance and Manufacturer’s Declaration –
Electromagnetic Emission
The Harmony
®
LED385 Examination Lighting System is intended for use in the electromagnetic environment
specified below. The Customer or the user of the Harmony LED385 Examination Lighting System should assure
that it is used in such an environment.
Emission Test Compliance Electromagnetic Environment –
Guidance
RF Emission
CISPR 11
Group 1 The Harmony LED385 Examination
Lighting System uses RF energy only
for its internal function. Therefore, its
RF emission is very low and is not likely
to cause any interference in nearby
electronic equipment.
RF Emission
CISPR 11
Class A The Harmony LED385 Examination
Lighting System is suitable for use in
all establishments other than domestic,
and may be used in domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes,
provided the following warning is
heeded:
Warning: This equipment/system is
intended for use by healthcare
professionals only. This equipment/
system may cause radio interference
or may disrupt the operation of nearby
equipment. If interference becomes a
concern, it may be necessary to
reorient or relocate the Harmony
LED385 Examination Lighting System
or shielding the location.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Compliance
EMC COMPLIANCE – TECHNICAL DATA
8