Operation of the Processor Operator Manual T6800 Rev B
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
7-17
Problem/Printout Probable Cause and Remedial Action
CONCENTRATION LOW
1. Check sterilant container:
Fresh sterilant container used?
Is there powder in the lower section of the cup?
Is the "pop-out" plug still in place?
Check the tray:
Is the screen at the base of the sterilant container well obstructed?
Check the Processor:
Did the chamber fill with water?
Did the pumps activate?
If the upper cup is not empty or powder is present in the lower part of the
cup, refer to the package insert for S40 Sterilant Concentrate or Section 3
of this manual and follow the instructions for disposing of a container of
sterilant, including the use of Personal Protective Equipment (PPE).
2. Replace container of sterilant and repeat cycle. If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
CONCENTRATION LOW
SEE OPERATOR MANUAL
CHAMBER LEVEL LOW
1. Lift processing tray out of processing chamber. If there is fluid in the drip pan
then dry it out – replace tray.
2. Repeat cycle. If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
CHAMBER LEVEL LOW
SEE OPERATOR MANUAL
TEMPERATURE OVER 60 DEG.
1. Check screen in sterilant compartment:
Remove any lint or debris.
Check inlet water temperature:
If over 60°C call your engineering department to have it adjusted.
2. Repeat cycle. If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
TEMP OVER 60 DEG
SEE OPERATOR MANUAL
POWER FAILURE
1. Check printer:
When paper feed button is pressed – does paper advance?
Was the Processor inadvertently unplugged?
Check all circuit breakers:
Processor, wall GFCI, and departmental.
2. Repeat cycle. If unsuccessful:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
POWER FAILURE
SEE OPERATOR MANUAL
TEMPERATURE DIFFERENCE TC1/TC2
1. Check screen in sterilant compartment:
Remove any lint or debris.
2. Repeat cycle. If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
TEMP DIFF TC1 TC2
SEE OPERATOR MANUAL
TROUBLESHOOTING A LIQUID CHEMICAL STERILIZATION CYCLE PROBLEM