T6800 Rev B Operator Manual Operation of the Processor
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
7-18
Problem/Printout Probable Cause and Remedial Action
HP PUMP FAILURE
1. Repeat cycle. If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
HP PUMP FAILURE:PUMP OFF/ON
SEE OPERATOR MANUAL
TEMPERATURE PROBLEM
1. Check room temperature:
Is it very cold (below 13°C/ 55°F)?
Is it very hot (above 65°C/149° F)?
2. Repeat cycle.
If same message appears:
CALL CUSTOMER SERVICE
WARNING
CYCLE
NOT COMPLETE
TEMP. PROBLEM
SEE OPERATOR MANUAL
CHAMBER OPEN FAULT
1. Check device and processing tray/container positioning.
Check for material and/or devices between top of processing tray/container and
Processor lid.
2. Close Processor lid and repeat cycle.
WARNING
CYCLE
NOT COMPLETE
CHAMBER OPEN
SEE OPERATOR MANUAL
TROUBLESHOOTING A LIQUID CHEMICAL STERILIZATION CYCLE PROBLEM