T6800 Rev B Operator Manual Quality Assurance
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
8-4
____________________________
DATE: 11-11-17
CYCLE START: 10:16:09A
REMARKS: ___________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
CYCLE
NOT COMPLETE
OPERATOR CANCELLED
SEE OPERATOR MANUAL
____________________________
____________________________
TEMP: 44 - 44 DEG. C
CONCENTRATION: 202
EXPOSURE TIME: 0.0 MIN
FILL TIME: 0.9 MIN
INLET TEMP: 38 DEG. C
:
OPERATOR ID: _______________
SERIAL # 0000000000000
CYCLE COUNT 10
CYCLE
CANCELLED: 10:18:13A
CHAMBER OPENED: 10:18:24A
OPERATOR ID: 12345678901245
DEVICE ID: 12345
PROCEDURE ID: 12
PATIENT: 1234567890123456
PHYSCN: 1234567890123456
WARNING
ESTIMATED
DAYS LEFT FOR
MAXLIFE FILTER: 13.4
SYSTEM 1 endo
DATE: Current date.
CYCLE START: Time cycle started.
OPERATOR ID: Space for the operator’s name or initials.
DEVICE ID: Record your specific device or load identification.
PROCEDURE ID: Record your specific procedure identification.
PATIENT: Record your specific patient’s identification.
PHYSCN: Record your specific physician’s identification.
REMARKS: Can be used to document reason and corrective action.
MESSAGE:
WARNING
CYCLE
NOT COMPLETE
OPERATOR CANCELLED
SEE OPERATOR MANUAL
TEMP: Documents the actual temperature range.
CONCENTRATION: Measures the actual equivalent units of conductivity of the
use dilution, not the concentration of peracetic acid.
EXPOSURE TIME: Documents the actual exposure time of the load to the use
dilution. Depending on when the cycle was cancelled, this may be less than the
6 minute exposure time necessary for liquid chemical sterilization.
FILL TIME: Number of minutes taken by the longest fill cycle.
INLET TEMP: Temperature of the water just before the heater is turned on after
the chamber is initially filled.
ESTIMATED DAYS LEFT FOR MAXLIFE FILTER: Number of days until the
MaxLife water filter will expire. This message will not appear until there are 15
or fewer days left.
OPERATOR ID: Space for the operator’s name or initials.
SERIAL #: Serial number of the Processor.
CYCLE COUNT: Current cycle count. This count does not include DIAGNOSTIC
cycles.
CYCLE CANCELLED: Time the cycle is cancelled.
CHAMBER OPENED: Time the chamber seal is deflated.
PRINTOUT PROVIDED FOR CANCELLED LIQUID CHEMICAL
STERILIZATION CYCLE INITIATED BY THE OPERATOR