Quality Assurance Operator Manual T6800 Rev B
SYSTEM 1 endo Liquid Chemical Sterilant Processing System
8-3
____________________________
DATE: 11-11-17
CYCLE START: 10:16:09A
REMARKS: ___________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
CYCLE
NOT COMPLETE
HEAT PROBLEM
SEE OPERATOR MANUAL
____________________________
____________________________
TEMP: 41.9 - 42.2 DEG. C
CONCENTRATION: 222
EXPOSURE TIME: 0.0 MIN
FILL TIME: 1.3 MIN
INLET TEMP: 45 DEG. C
: 13
OPERATOR ID: _______________
SERIAL # 0000000000000
CYCLE COUNT
CYCLE CANCELLED: 10:33:52A
CHAMBER OPENED: 10:34:32A
OPERATOR ID: 12345678901245
DEVICE ID: 12345
PROCEDURE ID: 12
PATIENT: 1234567890123456
PHYSCN: 1234567890123456
WARNING
ESTIMATED
DAYS LEFT FOR
MAXLIFE FILTER: 13.4
SYSTEM 1 endo
DATE: Current date.
CYCLE START: Time cycle started.
OPERATOR ID: Space for the operator’s name or initials.
DEVICE ID: Record your specific device or load identification.
PROCEDURE ID: Record your specific procedure identification.
PATIENT: Record your specific patient’s identification.
PHYSCN: Record your specific physician’s identification.
REMARKS: Can be used to document the corrective action.
MESSAGE:
WARNING
CYCLE
NOT COMPLETE
SPECIFIC PROBLEM IS DOCUMENTED
SEE OPERATOR MANUAL
TEMP: Documents the actual temperature. Depending on the reason for the
cancellation this may be different from the temperature necessary for liquid
chemical sterilization.
CONCENTRATION: Failure to use a new container of sterilant may result in a low
concentration reading causing the cycle to cancel. Concentration measures the
actual equivalent units of conductivity of the use dilution, not the concentration
of peracetic acid.
EXPOSURE TIME: Documents the actual exposure time of the load to the use
dilution. Depending on the reason for the cancellation, this may be less than the
6 minute exposure time necessary for liquid chemical sterilization.
FILL TIME: Number of minutes taken by the longest fill cycle. This time will
gradually increase with filter use. At 4.5 minutes, a notice will appear on the
printouts for completed cycles as a reminder to have a replacement MaxLife
water filter on hand. At 5 minutes, the cycle will cancel.
INLET TEMP: Temperature of the water just before the heater is turned on after
the chamber is initially filled. Or, if the cycle is cancelled first, it is the temperature
of the water dumped to drain before filling starts.
ESTIMATED DAYS LEFT FOR MAXLIFE FILTER: Number of days until the
MaxLife water filter will expire. This message will not appear until there are 15
or fewer days left.
OPERATOR ID: Space for the operator’s name or initials.
SERIAL #: Serial number of the Processor.
CYCLE COUNT: Current cycle count. This count does not include DIAGNOSTIC
cycles.
CYCLE CANCELLED: Time the cycle is cancelled.
CHAMBER OPENED: Time the chamber seal is deflated.
PRINTOUT PROVIDED FOR CANCELLED LIQUID CHEMICAL
STERILIZATION CYCLE INITIATED BY THE Processor