LaboPol-20
Instruction Manual
Original Instructions
Doc. no.: 16327025-02_A_en-us
Date of release: 2023.03.02

Table of Contents

Questions and Answers

  • A
    abigailtaylorAug 8, 2025
    What to do if Struers Laboratory Equipment is not draining water?
    • R
      Reginald AdamsAug 8, 2025
      If water isn't draining, check for these possible causes: * A squeezed drain hose: Straighten the hose. * A clogged drain hose: Clean the hose. * A drain hose that doesn't slope downwards: Adjust the hose to an even slope.
  • T
    Tyler StephensAug 12, 2025
    What to do if Struers Laboratory Equipment cooling water stops?
    • J
      jacqueline24Aug 12, 2025
      If the cooling water stops, check if: 1. The water tap on the water supply is closed: Turn the water on. 2. The built-in water tap is closed: Turn the water on. 3. The built-in water tap is blocked: Clean water tap. 4. The filter at the water inlet is blocked: Clean the filter with compressed air only.
  • W
    walterperezAug 16, 2025
    What to do if Struers Laboratory Equipment does not operate when the start switch is pressed?
    • M
      Michael WardAug 16, 2025
      If the machine doesn't start when you press the start switch: * Ensure that the main switch is on. Turn it on if it's off. * Check the fuse at the back of the machine and replace it if it's blown.
  • S
    Shawn Cruz MDAug 21, 2025
    What to do if Struers Laboratory Equipment has water dripping underneath?
    • K
      Kelli GarciaAug 21, 2025
      If you notice water dripping underneath the machine, it could be due to a leak in the water hose or a defect in the solenoid valve. To address this, turn off the main switch, disconnect the unit from the electrical power supply, turn off the water supply, and disconnect the unit from the water supply if needed. Contact Struers Service.
  • C
    Christine SaundersAug 26, 2025
    Why is there noise when Struers Laboratory Equipment starts, or why won't the turntable turn?
    • D
      Donna BaileyAug 26, 2025
      If you hear noise when the machine starts or the turntable won't turn, the belt might not be tight enough. The belt must be tightened. Contact Struers Service.
  • A
    Adam GravesAug 29, 2025
    What to do if Struers Laboratory Equipment specimen holder plate vibrates?
    • T
      Tanya MillerAug 29, 2025
      If the specimen holder plate vibrates, it may be due to: * Loose specimen holder plate screws: Tighten the screws. * An unbalanced specimen holder plate: Replace the specimen holder plate.
  • K
    Kelly McclureSep 2, 2025
    How to fix Struers Laboratory Equipment if the specimen mover head does not rotate?
    • H
      hgarciaSep 2, 2025
      If the specimen mover head does not rotate, check if the switch is set to “off”. If rotation is required, set the switch to “on.

Struers LaboPol-20 Specifications

General IconGeneral
TypeGrinding and Polishing Machine
Speed Range50-600 rpm
Frequency50/60 Hz
Weight25 kg
Disc/Wheel Diameter200 mm
Voltage230 V