9.4 Monthly
9.4.1 LaboForce-50 - pressure feet
The force on the pressure feet is generated by friction pins held in place by the screws in the
spring housing.
Use an Allen key to tighten the screws.
A Screws
9.5 Annually
9.5.1 Test the safety devices
The safety devices must be tested at least once a year.
WARNING
Do not use the machine with defective safety devices.
Contact Struers Service.
WARNING
Safety critical components must be replaced after a maximum lifetime of 20 years.
Contact Struers Service.
Note
Testing should always be performed by a qualified technician (electromechanical,
electronic, mechanical, pneumatic, etc.).
9 Maintenance and service