What to do if Struers AbraPol-30 Laboratory Equipment has water leakage?
V
Vanessa FloresAug 6, 2025
If you notice water leakage from your Struers Laboratory Equipment, it may be due to a leak in the cooling unit water hose. Inspect the hose for any signs of leakage and tighten the hose clamp to secure it.
S
Stacey FergusonAug 12, 2025
Why is water not draining away from my Struers AbraPol-30?
A
Amber MedinaAug 12, 2025
There are several reasons why water might not be draining from your Struers Laboratory Equipment:
* The drain hose may be squeezed, in which case you should straighten the hose.
* The drain hose could be clogged, so try cleaning it.
* The drain hose might not have a sufficient downward slope; adjust it to ensure an even slope.
S
samantha14Aug 17, 2025
What to do if specimens or equipment is corroded in Struers AbraPol-30 Laboratory Equipment?
A
Angela JonesAug 18, 2025
If specimens, the cooling unit, or other equipment components are corroded, it could be due to insufficient additive in the cooling fluid. To resolve this, add Struers cooling fluid additive to the cooling fluid, ensuring you use the correct concentration. A refractometer can be used to verify the concentration.
K
kimraymondAug 25, 2025
How to fix unclear display text on Struers AbraPol-30?
K
kevin67Aug 25, 2025
If the display text on your Struers Laboratory Equipment is unclear, it might be because the display is sensitive to temperature changes. You can adjust the brightness in the Configuration menu to improve clarity.
S
shawnwilliamsAug 30, 2025
Why is there continuous, irregular wear on the grinding/polishing surface of my Struers AbraPol-30?
K
Kevin MartinAug 30, 2025
Continuous, irregular wear on a grinding or polishing surface of your Struers Laboratory Equipment may be caused by a worn coupling on the specimen holder/mover plate or the specimen mover head. Replace the coupling. Contact Struers Service.