Service check
Struers offers a range of comprehensive maintenance plans to suit the requirements of our
customers. This range of services is called ServiceGuard.
The maintenance plans include equipment inspection, replacement of wear parts,
adjustments/calibration for optimal operation, and a final functional test.
7.10 Disposal
Equipment marked with a WEEE symbol contains electrical and electronic
components and must not be disposed of as general waste.
Contact your local authorities for information on the correct method of disposal in
accordance with national legislation.
For disposal of consumables and recirculation fluid, follow local regulations.
8 Troubleshooting - AbraPol-30
Error Cause Action
Display text is not clear. The display is sensitive to
temperature changes.
Change the brightness in the
Configuration menu.
Water is not draining away. Drain hose squeezed. Straighten the hose.
Drain hose clogged. Clean the hose.
Drain hose does not slope
downwards.
Adjust the hose to an even slope.
Continuous, irregular wear on
a grinding/polishing surface.
Coupling on the specimen
holder/mover plate or the
specimen mover head is
worn.
Replace the coupling.
Contact Struers Service.
Water leakage. A leak in the cooling unit water
hose.
Check the hose for leakages and
tighten the hose clamp.
Specimens, cooling unit or
equipment is corroded.
Not enough additive for
cooling fluid.
Add Struers cooling fluid additive
to the cooling fluid. Use the
correct concentration. Use a
refractometer to check. See the
Maintenance section.
8.1 Messages and errors - AbraPol-30
Error messages are divided into two classes:
Messages and errors
48 AbraPol-30
8 Troubleshooting - AbraPol-30

Table of Contents