Cleaning selected tubes
1. One by one, select the bottles you wish to clean. The status in the Select column
must be set to Yes
2. Press F1 to start the cleaning process. A pop-up message appears.
3. Follow the instructions in the pop-up message.
When the cleaning process is finished, the status in the Status column for the
clean bottle will change to Clean.
Cleaning used tubes
All tubes in use must be cleaned from regularly.
1. In the Cleaning of tubes screen, press F4 to select all used tubes.
2. To start the cleaning process, press F1.
3. Follow the instructions in the pop-up message.
When tube cleaning is finished, the status in the Status column for all bottles will
change to Clean.
7.4 Monthly
Recirculation unit
See the manual supplied with the specific equipment.
• Clean the recirculation unit.
• Replace the cooling fluid at least once a month.
Note
Replace the cooling fluid immediately if you notice that it is infected by algae or
bacteria.
7.4.1 Cleaning the recirculation unit
1. Clean the recirculation tank and the connected tubes thoroughly.
2. If you use a soap solution to clean the bowl or the recirculation tank, rinse with clean water
before filling the recirculation tank.
Note
If the cooling fluid is contaminated by algae or bacteria, replace the cooling fluid
immediately.
3. If the cooling water has been infected with bacteria or algae, clean the tank and tubes with a
suitable antibacterial disinfectant.
4. Clean the static filter: Remove it and rinse it with water.