Safety Related Parts of the Control System (SRP/CS)
For specific safety related parts, see the section "Safety Related Parts of the Control System
(SRP/CS)" in the section "Technical data" in this manual.
Technical questions and spare parts
If you have technical questions or when you order spare parts, state serial number and
voltage/frequency. The serial number and the voltage are stated on the type plate of the machine.
7.1 General cleaning
To ensure a longer lifetime for your machine, Struers strongly recommends regular cleaning.
Note
Do not use a dry cloth as the surfaces are not scratch resistant.
Note
Do not use acetone, benzol or similar solvents.
If the machine is not to be used for a longer period of time
Clean the machine and all accessories thoroughly.
7.2 Daily
Clean all accessible surfaces with a soft, damp cloth.
Recirculation unit
See the manual supplied with the specific equipment.
Check the level of the cooling fluid after 8 hours of use, or at least once a week. If needed, fill up
the tank with more cooling fluid.
Check the filters. Clean the filters, if needed.
7.2.1 Checking the recirculation tank
1. Check and, if necessary, change the cooling fluid.
Note
If the cooling fluid is contaminated by algae or bacteria, replace the cooling fluid
immediately.
2. If the pump cannot reach the cooling fluid, refill the unit.
3. Change the cooling fluid if it is dirty. See Changing the cooling fluid 43.