Replacing the safety cover screen
Note
The safety cover screen must be replaced immediately if it has been weakened by
collision with projectile objects or if there are visible signs of deterioration or damage.
Note
The screen must be replaced to remain compliant with the safety requirements stated
in EN 16089.
To ensure its intended safety, the safety cover screen must be
replaced every 5 years. A label on the screen indicates when it
must be replaced.
Recirculation unit
See the manual supplied with the specific equipment.
7.5.1 Testing safety devices
The safety devices must be tested at least once a year.
WARNING
Do not use the machine with defective safety devices.
Contact Struers Service.
Note
Testing should always be performed by a qualified technician (electromechanical,
electronic, mechanical, pneumatic, etc.).
See Emergency stop ►44.
See Safety cover ►45.
7.5.2 Emergency stop
Test 1
1. Press the Start button. The machine starts operating.
2. Press the emergency stop.
3. If operation does not stop, press the Stop button.
4. Contact Struers Service.