1. Open the electrical connection box.A
2. Connect a 4-lead or 5-lead cable as shown.
PE Earth (ground)
N Neutral (not used internally)
L1 Phase
L2 Phase
L3 Phase
EU cable
L1 Brown
L2 Black
L3 Black or Grey
Earth (ground) Yellow/Green
Neutral Blue
UL cable
L1 Black
L2 Red
L3 Orange/Turquoise
Earth (ground) Green (or Yellow/Green)
Neutral White
The other end of the cable can be fitted with an approved plug or hard-wired into the power supply
according to the electrical specifications and local regulations.
5.4 Noise
For information on the sound pressure level value, see this section: Technical data ►54,
CAUTION
Prolonged exposure to loud noises may cause permanent damage to a person’s
hearing.
Use hearing protection if the exposure to noise exceeds the levels set by local
regulations.
5.5 Vibration
For information on the total vibration exposure to hand and arm, see this section: Technical data
►54.
5 Installation
AbraPol-30 15