What to do if Struers LaboPol-60 Laboratory Equipment machine does not operate when the start switch is pressed?
D
Donna Caldwell DDSAug 16, 2025
If your Struers Laboratory Equipment does not start when you press the start switch, there are a couple of things to check:
* First, make sure the main switch is turned on.
* Second, check the fuse located at the back of the machine and replace it if it's blown.
E
edwardscameronAug 19, 2025
Why is my Struers LaboPol-60 Laboratory Equipment cooling water stopping?
K
Kayla GarciaSep 12, 2025
If the cooling water stops in your Struers Laboratory Equipment, consider these potential causes:
1. The water tap on the water supply might be closed. Ensure it is turned on.
2. The built-in water tap could be closed. Open it to allow water flow.
3. The built-in water tap might be blocked. Clean it to remove any blockage.
4. The filter at the water inlet could be blocked. Clean the filter using compressed air.
U
uwashingtonAug 22, 2025
What to do if Struers Laboratory Equipment water is not draining away?
L
lynndustinAug 22, 2025
If water isn't draining from your Struers Laboratory Equipment, check the following:
* Make sure the drain hose isn't squeezed. Straighten it out if it is.
* Check if the drain hose is clogged and clean it if necessary.
* Ensure the drain hose slopes downwards evenly; adjust it if needed.
A
Amy WardAug 25, 2025
What to do if Struers Laboratory Equipment preparation disc runs unevenly or stops?
E
erincervantesAug 26, 2025
If the preparation disc on your Struers Laboratory Equipment runs unevenly or stops, it might be due to excessive force. Try reducing the force being applied.
J
jonesryanAug 27, 2025
Why is there noise when Struers Laboratory Equipment machine starts, or the turntable will not turn?
A
Angela NormanAug 27, 2025
If you notice noise when starting your Struers Laboratory Equipment, or if the turntable isn't turning, the belt might not be tight enough. The belt must be tightened. Contact Struers Service.
T
Tracy RamosAug 30, 2025
What to do if Struers Laboratory Equipment water is dripping underneath the machine?
C
Christopher WilsonAug 30, 2025
If you find water dripping underneath your Struers Laboratory Equipment, it could be due to a leak in the water hose or a defect in the solenoid valve. Turn the main switch off. Disconnect the unit from the electrical power supply. Turn the water supply off. Disconnect the unit from the water supply, if needed. Contact Struers Service.
J
Joseph CooperSep 1, 2025
What to do if Struers LaboPol-60 Laboratory Equipment preparation disc stops?
J
Jill RogersSep 2, 2025
If the preparation disc stops on your Struers Laboratory Equipment, the frequency inverter may have stopped the equipment. Switch the equipment off, wait a few minutes, and then restart it. If the error persists: Contact Struers Service.
T
Troy SanchezSep 4, 2025
What to do if Struers LaboPol-60 Laboratory Equipment specimen holder plate vibrates?
F
Faith MartinSep 4, 2025
If the specimen holder plate on your Struers Laboratory Equipment vibrates, it could be due to:
* Loose specimen holder plate screws. Tighten them.
* An unbalanced specimen holder plate. Replace it.
M
Mary SheppardSep 7, 2025
What causes the Struers LaboPol-60 specimen mover to start turning?
D
David ClarkeSep 8, 2025
If the specimen mover starts turning on your Struers Laboratory Equipment, it could be due to:
* Loose column screws. Tighten the screws immediately.
* The column not being securely mounted. Turn the column until the spacer sits securely in the support hole.
T
Timothy MontgomerySep 10, 2025
Why is my Struers Laboratory Equipment giving bad preparation results from the same method?
L
larry69Sep 12, 2025
If you're getting poor preparation results from your Struers Laboratory Equipment using the same method, it might be because the pressure foot turns by itself, leading to differing forces. Increase friction by tightening the screws in the housing. See the section "Maintenance."