Water supply specifications
Water pressure 1 to 9.9 bar (14.5 to 143 psi)1 - 9.9 bar (14.5 - 143 psi)
Hose supplied Diameter: ¾”. Length: 2 m (7’).
With standard connector, filter gasket and reduction ring (¾" to
½").
Tube connection ¾" British Standard pipe thread.
Water outlet - drain
Note
Make sure that the waste water hose slopes downwards towards the waste water
drain throughout its length.
Water outlet specifications
Hose supplied Diameter: 40 mm (1.6"). Length: 1.5 m (4.9’)
15.13 Compressed air
Note
Required for LaboForce-100 (optional).
Compressed air specifications
Pressure 6 - 9.9 bar/87 - 143 psi
Air consumption, approx. 3.5-4 l/min
Air quality Class-3, as specified in ISO 8573-1
Hose supplied 6 mm inlet tube (blue) with quick coupling connector
15.14 Recirculation cooling unit
The recirculation cooling unit is optional.
Recommended
• Struers Cooling System 3
• Struers additive for recirculation cooling unit to prevent corrosion.
Recirculation Cooling Unit specifications
Power supply 2.5 m (8.2') electrical power cable for connection to a single-
phase electrical power supply.
Water supply 2.5 m (8.2') water hose and GEKA connection.
110 LaboPol-60
15 Pre-installation checklist